4-Quardrant Operation

A 4-quadrant power supply can operate in all four voltage-current combinations, enabling it to both source and sink power across positive and negative polarities, which is essential for testing bidirectional and regenerative systems.

AC DC Programmable power supplies by actionpower test ATE

What is Four-Quadrant Operation?

4-Quadrant operation is typically specified capability of a power supply that is able to source and sink voltage and current both positively and negatively. To explicate in a way more intelligible combined with the I-V plane where the horizontal axis stands for the current and the vertical stands for the voltage, a four-quadrant power supply can operate covering all the Cartesian coordinate system’s 4 quadrants.

There is also frequency ambivalence between terminologies including bipolar power supply, bidirectional power supply, 2-quadrant power supply, apart from the 4-quadrant operation power supply. The below are detailed explanations, compare and contrast to address their differences.

Distinguish Source and Sink in the Four-Quadrant I-V Plane

To spot whether each of the four quadrants stands for that the power supply is sourcing or sinking power, one may count on the Ohm’s law of electrical power to tell. In a short word, they unit sinks power in case the power calculated is negative, and vice versa in case of sourcing power.

Sink and Source Compared

  • Source if \(P > 0\): the device delivers energy to the outside.
  • Sink if \(P < 0\): the device absorbs energy from the outside.

In detail, in a four-quadrant I–V coordinate system,

  • the horizontal axis is voltage \(V\) (polarity),
  • the vertical axis is current \(I\) (direction),
  • and instantaneous power is \(P = V \times I\)

This is simply a mathematical expression of the energy-flow direction.

Correspondingly in Each of the 4 Quadrants in the System

QuadrantVoltage \(V\)Current \(I\)Power \(P = V I\)Interpretation
I\(+\)\(+\)\(+\)Forward Source (conventional supply)
II\(-\)\(+\)\(-\)Sink (reverse voltage, absorbing current)
III\(-\)\(-\)\(+\)Reverse Source (supplying with reversed polarity)
IV\(+\)\(-\)\(-\)Sink (positive voltage, regenerative current)

Uni-Polar, Bi-Polar, 4 Quadrant Operation Compared and Explained

Source: Power is positive. Current and voltage are of the same direction.

Sink: Power is negative. Current and voltage are of opposite direction.

Uni-Polar Power Supply with Operation Capability in 1st Quadrant Only

A uni-polar power supply operates only within the 1st quadrant of the coordinate system and restricted to running by supplying positive current and positive voltage only, or alternatively within the device \(I > 0\), \(V > 0\), \(P > 0\). One of the typical examples among our category is the PDC laboratory DC power supply that is mere a programmable source of DC power.

Negative Polarity Power Supply with Operation Capability in 3rd Quadrant Only

When a power supply performs negative polarity output where \(I < 0\), \(V < 0\), \(P > 0\), the unit is still sourcing power while with reversed polarity compared to the first quadrant. The application of such kind power supply is for testing scenarios that the DUT confronts reverse-polarity supply, examples including:

  • Devices requiring bipolar power (e.g., operational amplifiers, motor drivers).
  • Simulating risks of reverse charging in batteries.
  • Reverse-polarity experiments in motor control and inverter testing.

Voltage-current coordinate system diagram. The vertical axis represents Voltage (+V, -V) and the horizontal axis represents Current (+I, -I). The top-right first quadrant is highlighted in blue and labeled 'QUADRANT I SOURCE (+V, +I)', indicating the device operates as a power source outputting positive voltage and positive current.

Bidirectional Power Supply or Electronic Load Working in 2nd Quadrant to Sink Power

A bidirectional power supply or electronic load operating in the 2nd quadrant is defined by the conditions \(V < 0\), \(I > 0\), and therefore \(P = V \times I < 0\).

In this mode, the device acts purely as a sink, absorbing energy from the DUT under a negative voltage while current flows into the instrument. Typical use cases include battery charging simulations, testing negative voltage rails, or evaluating reverse-polarity operating conditions.

The absorbed energy may be internally dissipated as heat in conventional designs, or, in the case of regenerative systems, it can be efficiently fed back to the grid.

Two side-by-side voltage-current quadrant graphs. The vertical axis is Voltage and the horizontal is Current. The left graph highlights only the top-left Quadrant II in orange, labeled 'QUADRANT II SINK (+V, -I)'. The right graph shows the same sink quadrant plus the top-right Quadrant I highlighted in blue, labeled 'QUADRANT I SOURCE (+V, +I)'.

 

Four-Quadrant Operation Power Supply

A four-quadrant operation power supply can operate in all four regions of the I–V plane, sourcing and sinking current under both positive and negative voltage conditions. In other words, it handles energy flow in either direction and supports both polarities: it can act as a source or as a load, with \(V\) and \(I\) of either sign. Instantaneous power is \(P = V \times I\). Typical applications of four-quadrant power supplies may include:

A graph illustrating four-quadrant operation on a voltage-current plane. The vertical axis represents Voltage (+V up, -V down) and the horizontal axis represents Current (-I left, +I right). Quadrants I and III are colored blue and labeled as 'Source' with (+V, +I) and (-V, -I) respectively. Quadrants II and IV are colored orange and labeled as 'Sink' with (+V, -I) and (-V, +I) respectively.

  • Battery simulation: discharge as a source (Quadrant I), charge as a sink (Quadrant II).
  • Motor drive / inverter testing: motoring as a source, braking energy returned as a sink (Quadrant IV).
  • Bipolar analogue systems: \(\pm V\) rails for op-amps and mixed-signal circuits (Quadrants I & III).
  • Grid-tied devices: exercising power flow in and out under realistic line conditions.
QuadrantVoltage \(V\)Current \(I\)Power \(P = V I\)Function
I\(V > 0\)\(I > 0\)\(P > 0\)Source power under positive voltage (conventional supply)
II\(V < 0\)\(I > 0\)\(P < 0\)Sink power under negative voltage (e.g., battery charging, negative rail testing)
III\(V < 0\)\(I < 0\)\(P > 0\)Source power under negative voltage (reverse-polarity supply)
IV\(V > 0\)\(I < 0\)\(P < 0\)Sink power under positive voltage (regenerative testing, current sinking)

Regeneration (if equipped): when \(P<0\), absorbed energy may be dissipated as heat or fed back to the grid in regenerative models.

Note: Here “regenerative” refers to the power supply’s ability to process sink energy. It is not the same as a dedicated regenerative electronic load, which is a specific instrument type designed to return absorbed energy to the grid. A four-quadrant regenerative AC power supply is usually referred to as a grid simulator.

Quick Model Access

Smart Discovery

Access our Power Supply Wizard to find the right programmable power supply for your use cases and testing scenarios.

discover programmable power supply by filtering with current, voltage and power ranges