Immunity Testing According to IEC 61000-4-13: Test Principles, Scenarios, and Programmable Power
This application note provides a technical overview of IEC 61000-4-13, a standard established to evaluate the immunity of grid-connected equipment against interharmonic voltage disturbances commonly encountered in modern power systems.
Building upon the previous application note on IEC 61000-4-7, which introduces the fundamentals of harmonics and interharmonics as well as their measurement methods, this session focuses on the practical test applications defined in IEC 61000-4-13 and their implementation in immunity testing.
In case that you are looking for programmable power for above mentioned testing applications, get in contact with ActionPower for professional suggestions about model options.
| Version | 1.0.1 |
|---|---|
| Categories | Harmonics |
| Download | Download Application Note |
| Download Count | 12 |
| Page Views | 2492 |
| Size | 964.37 KB |
| Create Date | 23/01/2026 |
| Last Updated | 04/11/2026 |
| Author | Zhiwei Bao |
| Reviewed | Sen Wang |
Power Quality as a Persistent Challenge for Interconnected Power Electronic Equipment
Modern power systems are no longer characterized by purely sinusoidal voltage waveforms. As power electronic converters become increasingly interconnected, power quality disturbances – particularly interharmonics – have emerged as a persistent challenge rather than an exception.
Resulting from non-linear loads whose operations are characterized by non-synchronous switching processes, frequency conversions, and, modulation strategies, for example, the SMPS [1], inter-harmonics represents a particularly challenging class of disturbances that significantly affect and influence control stability, measurement accuracy, and protection behaviour of grid-connected equipment. Moreover, these components often fall outside the frequency ranges traditionally addressed in harmonic analysis and therefore require dedicated measurement approaches.
Devices affected by interharmonics, while not necessarily experiencing immediate functional failure, may nevertheless exhibit subtle yet critical effects, such as oscillatory control responses, unexpected protection triggering, increased thermal stress, or audible noise. These behaviour are highly dependent on interharmonic spectral parameters yet they are difficult to predict without dedicated immunity testing.

Fig 1. Grid Inter-Connection is more Complex than Ever
Consequently, IEC 61000-4-13 [2] establishes standardized test procedures to assess the response of equipment to interharmonic voltage disturbances in a controlled and reproducible manner.
IEC 61000-4-13 is part of (EMC) Directive 2014/30/EU, against which manufacturers in the electronics or electrical industries should have their products in compliance with. Discover what is EMC Directive from this article [3]
Overview of Test Defined in IEC 61000-4-13: Test Scenarios and Implementation
IEC 61000-4-13 establishes a common reference for evaluating the functional immunity of electrical and electronic equipment at the AC power port when subjected to harmonic and interharmonic voltage disturbances (including mains signalling) on low-voltage power networks.
The standard specifies a set of voltage distortion conditions to represent modern public supply environments influenced by non-linear and power-electronic loads, where harmonic/interharmonic components are intentionally superimposed on the fundamental mains frequency.

Fig 2. IEC 61000-4-13: Testing and Measurement Techniques – Harmonics and Interharmonics including Mains Signalling at A.C. Power Port, Low Frequency Immunity Tests
Two representative time-domain distortion waveforms are defined as reference profiles:
- Flat curve (flat-top): a waveform whose each half-wave is formed by a sine segment followed by a constant-voltage “flat” region and then a sine segment, with the resultant RMS maintained at nominal.
- Over swing: a waveform created by superimposing discrete harmonic components – typically the 3rd and 5th harmonics with specified phase relationships – to produce peak amplification behaviour
Apart from standardized fundamental components, IEC 61000-4-13 is restricted to electrical devices and equipment with current no more than 16 A per phase, while the disturbance frequencies superimposed extending only to 2kHz for 50 Hz systems (and proportionally 2.4kHz for 60 Hz systems), which are defined and intended for testing against DUTs to be used in residential, commercial, and light industrial environments.
Note: Equipment connected to special-purpose networks such as 16⅔ Hz railway [4] or 400 Hz aerospace systems is excluded. Refer to this article for relevant information.
Collectively, these test scenarios define the types of harmonic and interharmonic disturbances to be applied and the immunity aspects to be observed, without prescribing specific implementation techniques. The detailed execution of these scenarios and their practical realization are addressed through the individual test items discussed in the following section.
Key Immunity Test Items Defined in IEC 61000-4-13: Examples and Technical Introductions
IEC 61000-4-13 structures its immunity evaluation around a limited set of deliberately simplified, yet field-observed voltage distortion scenarios. Rather than attempting to reproduce arbitrary grid waveforms, the standard defines representative disturbance constructs that allow functional immunity to be assessed in a controlled and repeatable manner.
The standard formalizes three core disturbance dimensions:
- Spectral placement of harmonic and interharmonic components
- Amplitude severity relative to the fundamental voltage
- Temporal application rules, including dwell time and sweep rate
These dimensions jointly define the test items described below.
Voltage Synthesis Framework Defined by the Standard
All test items specified in IEC 61000-4-13 are based on a synthesized supply voltage in which harmonic and interharmonic components are intentionally superimposed onto the fundamental mains waveform, while the resultant RMS voltage is strictly maintained at nominal value [2] (Clause 5 – Test levels, EC 61000-4-13:2002+A1:2009+A2:2015) throughout the test.
$$
v(t) = V_1 \sin(\omega_1 t + \phi_1)
+ \sum_{n=2}^{N} V_n \sin(n \omega_1 t + \phi_n)
+ \sum_{k=1}^{M} V_{ih,k} \sin(\omega_{ih,k} t + \phi_{ih,k})
$$
where:
- v(t): instantaneous test voltage applied to the DUT
- V1: RMS (or peak, consistently defined in the test plan) magnitude of the fundamental component
- ω1 = 2πf1: angular frequency of the fundamental, with f1 ∈ {50, 60} Hz
- φ1: phase angle of the fundamental component
- Vn: magnitude of the n-th harmonic voltage component (integer multiple of f1)
- n: harmonic order index (n ∈ Z+)
- N: highest harmonic order included in the synthesized test waveform
- Vih,k: magnitude of the k-th interharmonic voltage component
- ωih,k = 2πfih,k: angular frequency of the interharmonic component, where fih,k ≠ n f1
- φih,k: phase angle of the interharmonic component
- M: number of interharmonic components superimposed
This constraint – constant RMS voltage under increasing spectral distortion – is central to the philosophy of IEC 61000-4-13 and distinguishes immunity testing from emission characterization.
Test Items Summary
| Test item (engineering view) | Applied disturbance | Primary controlled variables | Typical observation focus | Why it matters |
|---|---|---|---|---|
| Single interharmonic injection | One interharmonic component added on top of f1 | fih, Vih/V1, φih | Control oscillation, protection mis-trip, regulation deviation, audible noise | Isolates frequency-specific sensitivity and reveals coupling points |
| Swept interharmonic frequency | One interharmonic component swept across a band | [fstart, fstop], Δf / r, constant Vih | Resonance amplification, PLL loss-of-lock, reduced stability margin | Captures frequency-dependent immunity behaviour missed by fixed tests |
| Composite interharmonic disturbance | Multiple interharmonics present simultaneously | {fih,k}, {Vih,k}, {φih,k}, M | Nonlinear interaction, cumulative stress, mode transitions | Replicates spectral congestion and nonlinear superposition effects |
Single Interharmonic Voltage Injection
This test item represents the most fundamental form of interharmonic immunity evaluation defined in IEC 61000-4-13. By intentionally introducing a single interharmonic voltage component onto an otherwise nominal mains waveform, the objective is to isolate the interaction between a specific non-integer frequency component and the internal control, filtering, or protection mechanisms of the device under test.

Fig 3. IEC 61000-4-13 Test Conditions: Non-Integer Odd-Order Harmonics, Class 3 (50 Hz) with ActionPower CORTEX AC
Unlike composite or swept conditions, this test deliberately avoids spectral congestion. As a result, any observable deviation in behaviour can be directly associated with a single, well-defined spectral excitation.
$$
v(t) = V_1 \sin(\omega_1 t) + V_{ih} \sin(\omega_{ih} t + \phi_{ih}),
\quad f_{ih} \neq n f_1
$$
where:
- Vih: magnitude of the injected interharmonic component
- fih: interharmonic frequency (non-integer multiple of f1)
- φih: interharmonic phase angle relative to the fundamental reference
This test item serves as a baseline probe. By injecting a single interharmonic component, observed behaviour can be
attributed to a specific spectral excitation rather than cumulative effects.
Swept Interharmonic Frequency Test
While fixed-frequency excitation is effective for identifying isolated sensitivities, interharmonic disturbances in practical power systems are rarely stationary. Their frequency content may drift due to asynchronous operation of converters, control interaction, or changing grid conditions.

Fig 4. Interharmonics Frequency Sweeping with ActionPower TITAN Series Grid Simulator
To capture this behavior, IEC 61000-4-13 introduces swept interharmonic testing, where the disturbance frequency is systematically varied while the disturbance magnitude remains constant. This approach allows frequency-dependent immunity characteristics and resonance-related phenomena to be revealed.
$$ \text{Continuous sweep: }
f_{ih}(t) = f_{start} + r t
$$
$$ \text{Discrete-step sweep: }
f_{ih,i} = f_{start} + i \Delta f
$$
$$ \text{with constant disturbance magnitude: }
V_{ih} = \text{const.}
$$
where:
- fstart, fstop: sweep start and stop frequencies
- r: sweep rate (Hz/s)
- Δf: frequency step size
- i: step index
This test identifies frequency-dependent immunity characteristics and resonance-related sensitivity that cannot be captured by static excitation, particularly in systems with narrow control bandwidths or frequency-selective dynamics.
Composite Interharmonic Disturbance Test
Modern low-voltage power networks typically host multiple non-linear and power electronic loads operating asynchronously. Under such conditions, interharmonic disturbances rarely appear as isolated spectral components; instead, multiple interharmonics coexist and interact within the same supply environment.
This test item reflects that reality by applying several interharmonic voltage components simultaneously. The objective is not merely to increase disturbance severity, but to evaluate how the device under test responds to spectral superposition and potential nonlinear interaction effects.
$$
v(t) = V_1 \sin(\omega_1 t)
+ \sum_{k=1}^{M} V_{ih,k} \sin(\omega_{ih,k} t + \phi_{ih,k})
$$
where:
- M: number of interharmonic components
- Vih,k, fih,k, φih,k: magnitude, frequency, and phase of each component
This test reflects realistic grid conditions with multiple asynchronous converters and reveals nonlinear interaction effects that are not observable under single-frequency injection or swept excitation alone.
Test Setup for IEC 61000-4-13 Immunity Evaluation and Source Requirements
The standard defines test configurations to which the laboratory-based setup shoud typically refer to where a series of devices and measurement instruments are used for generating harmonics and interharmonics, voltage disturbances, and which in turn to be applied under controlled conditions. The mains supply is used only as an upstream energy source, while the actual test voltage applied to the equipment under test (EUT) is synthesized by a waveform generator (a remote control with testing software) and delivered through a controlled and programmable AC source.

Fig 5. Single-Phase and Three-Phase Test Set-Up Topologies as per IEC 61000-4-13
Single-Phase Test Setup
In a typical single-phase IEC 61000-4-13 test setup, the programmable AC power source acts as the controlled interface between the utility mains and the equipment under test (EUT). The mains input provides only the upstream energy supply, while the test waveform – including the fundamental component and any superimposed harmonic or interharmonic disturbances – is generated internally by the AC source under computer control.
The synthesized test voltage is applied to the EUT through line (L) and neutral (N), with protective earth (PE) connected in accordance with normal operating conditions. Voltage and current are measured at the interface between the source and the EUT, ensuring that the actual electrical stress experienced by the EUT is observed rather than the idealized source waveform alone.

Fig 6. Single-Phase Test Set-Up Example
This single-phase configuration represents the most common and fundamental test scenario defined in IEC 61000-4-13 and is typically used as the baseline for evaluating interharmonic immunity, control stability, and functional behavior under defined voltage distortion conditions.
Note on Three-Phase Configuration
For three-phase equipment, the same principle applies: three synchronized source channels supply L1, L2, and L3 with a common reference, while phase voltages and currents are measured at the EUT interface. The overall test philosophy and signal flow remain identical to the single-phase case.
Source Requirements as per IEC 61000-4-13
The key factor for a sucessful and compliance harware setup for IEC 61000-4-13 test bench lies with the selection of programmable AC power source that meets those requirements and specificatiosn listed in the standard in Chapter 6.1 and table 5. To be more specific, we compare those requirements listed there with the CORTEX AC Programmable Power Source by ActionPower.
CORTEX programmable AC Power Supply is a bidirectional and regenerative programmable power supply designed in modular and rack-mount architecture for grid-related testing from low to medium-level power application.
In case you are looking for a programmable AC source for relevant testing applications, visit the product page of CORTEX AC Power Source.
| Parameter | IEC 61000-4-13 Requirement (Source / Generator) | CORTEX Capability |
|---|---|---|
| Output current capability | Generator shall provide an output sufficient to test the EUT up to a maximum rated input current of 16 A r.m.s. per phase (unless product standard specifies otherwise). | CORTEX series provides much higher phase current capability depending on model. |
| Fundamental voltage magnitude (U1) | Nominal mains voltage ±2% (single-phase / three-phase). | Voltage accuracy: ±(0.01% + 0.05% F.S.) (0.001–400 Hz) |
| Fundamental frequency (f1) | 50 Hz ±0.5% or 60 Hz ±0.5%. |
|
| Angle between phases | 120° ±1.5° (star connection). | Phase angle control: range 0–359.9°; resolution 0.1°; accuracy ±0.1° (0.001–200 Hz). Supports single/three-phase and independent 3φ control. |
| Selectable individual harmonics (order) | 2 to 40. | Harmonic injection range: up to 100th order (at 40–70 Hz fundamental); up to 25th order (at 70–200 Hz fundamental). |
| Individual harmonic magnitude range | 0% to 14% of U1. | Harmonic injection supported; spec provides “Content 40%” and magnitude error definition. |
| Individual harmonic magnitude accuracy | The larger of ±5% of Uh or ±0.1% of U1. | Magnitude error: ±5% @ set value or 0.1% of the fundamental (same expression form as IEC requirement). |
| Harmonic phase angle (h = 2…9) | Phase angles: 0° or 180°; accuracy of zero phase crossing displacement with respect to fundamental: ±2°. | Phase angle programmable 0–359.9° with stated control accuracy ±0.1° (note: IEC only needs 0/180 plus displacement tolerance). |
| Interharmonic magnitude range | 0% to 10% of U1. | “Inter-harmonics” listed under transient functions (feature available). |
| Interharmonic magnitude accuracy | The larger of ±5% (of set) or ±0.1% of U1. | Harmonic magnitude error in IEC-like form; interharmonic. |
| Interharmonic frequency range | 0.33×f1 to 40×f1 (≈ 16.5–2000 Hz for 50 Hz; 19.8–2400 Hz for 60 Hz). |
|
| Interharmonic frequency step / adjustment | Frequency set rules by band (e.g., step sizes expressed as fractions of f1); maximum error of adjusted value ±0.5% of f. | Frequency resolution 0.001 Hz; frequency accuracy ±0.01% (significantly finer than IEC’s ±0.5% f allowance). |
| Ability to superimpose multiple components | Generator shall provide the option to superimpose more than one voltage on each phase. | Supports harmonic injection + inter-harmonics + programmable waveforms / DST; suitable for composite superposition scenarios. |
| Control of magnitude / frequency / phase & sequencing | Generator shall provide control inputs for selection of magnitude, frequency, phase-angle, and sequence of the superimposed voltage. | Computer control + programmable waveform lists / step / pulse / advanced / harmonics / inter-harmonics; phase angle control and independent 3φ control supported. |
| Output impedance / setpoint at EUT terminals | No fixed output impedance defined; internal source impedance must be controlled so voltage drop on internal/connection impedance is compensated and set values are met at EUT terminals (connections kept short). | Adaptive voltage compensation is specified; regulation specs provided (line/load regulation), consistent with maintaining setpoint at the EUT interface. |
| Verification / inherent distortion | Generator output should be verified (typically with IEC 61000-4-7 Class A measurement method); inherent distortion limits apply before adding disturbances. | Voltage distortion figures are specified by frequency band (e.g., low THD at 50/60 Hz; higher-frequency bands specified separately), supporting pre-test verification planning. |
Application Scenarios Where IEC 61000-4-13 Matters
As power electronics penetration increases, interharmonic immunity is no longer a niche requirement but a baseline robustness indicator. In real installations, interharmonics may originate from converter switching patterns, asynchronous control interactions, sub-synchronous oscillations, and network impedance resonances, often appearing as frequency components that are neither integer harmonics nor steady-state “clean” tones. For grid-tied equipment, the risk is rarely an immediate catastrophic failure; rather, it manifests as subtle but consequential behaviour degradation that only becomes visible under controlled excitation.
Typical scenarios where IEC 61000-4-13-oriented interharmonic immunity evaluation becomes relevant include:
Grid-tied inverters (PV, wind, BESS PCS, microgrid interfaces)
These systems synchronize to the grid through PLL and control loops that can be sensitive to interharmonic components near control bandwidth boundaries. Interharmonics may trigger oscillatory current injection, loss of synchronization margin, or protection mis-operation, particularly when multiple converters share a weak grid with resonant impedance.
Utility-scale grid-adaptability testing is usually conducted at high power reaching MVA-level, for higher power testing applications in relation to IEC 61000, refer to TITAN Series Grid Simulator.
In difference to traditional AC power load that comes in resistive form and dissipates energy, TITAN high power grid simulator is designed with regenerative technology to sink power and feed it back to grid with an efficiency rate of 91% to save your power testing cost.
EV onboard chargers and bidirectional OBCs (V2G/V2H/V2L-capable)
OBCs operate under wide operating envelopes and may interact with upstream supply distortion and charger-side converters. Interharmonics can lead to unexpected current ripple, audible noise, unstable control responses, or conservative derating modes that impact charging performance and efficiency.
Industrial motor drives and servo systems
Drives often integrate rectifier stages, DC link dynamics, and fast current control. Interharmonics in the supply can translate into torque ripple, acoustic noise, controller jitter, or nuisance trips—especially in environments where multiple drives and SMPS loads coexist.
Infrastructure and power conversion equipment in commercial/industrial LV networks
Data centers, factory automation lines, building power systems, and UPS/PDU ecosystems commonly exhibit spectral congestion. Interharmonic immunity becomes a practical reliability metric, reducing the likelihood of false alarms, control instability, and intermittent malfunction under distorted mains conditions.
Wrap-Up and Key Takeaways
Interharmonics are no longer exceptional events but a persistent feature of modern low-voltage grids with high penetration of power electronic converters. Their impact on grid-connected equipment is typically manifested as behavioral instability rather than immediate failure, making controlled immunity evaluation essential.
IEC 61000-4-13 establishes a practical framework for this purpose by defining repeatable interharmonic disturbance conditions, in terms of frequency placement, magnitude, phase relationship, and superposition, so that equipment response can be assessed under deterministic and reproducible conditions that cannot be achieved on a live grid.
Key takeaways include:
- Interharmonic immunity is a functional robustness issue, not a waveform compliance exercise.
- Controlled disturbance synthesis is essential to isolate frequency-dependent and interaction-driven effects.
- The test source’s capability directly determines whether IEC 61000-4-13 requirements can be realistically implemented.
ActionPower’s programmable AC power sources and grid simulation platforms provide the required control depth in frequency, amplitude, phase, and superposition. Get in contact with us in case you are looking for a programmable power source for controlled, repeatable, and engineering-relevant methods executing IEC 61000-4-13 interharmonic immunity testing.
Bibliography
- [1] What is IEC 61000-4-13: A Practical Guide for Manufacturers, Advanced Testing Eequipment Corp., Link
- [2] Electromagnetic compatibility (EMC) – Part 4-13: Testing and measurement techniques – Harmonics and interharmonics including mains signalling at a.c. power port, low frequency immunity tests, International Electrotechnical Commission, Link
- [3] Electromagnetic Compatibility (EMC) Directive, European Commission, Link
- [4] Overcoming barriers to the Single European Railway Area, Global Railway Review, Link

Zhiwei Bao
Zhiwei Bao holds 20+ patents in the field of mechanical engineering and is keen on industrial automation and product marketing. His interdisciplinary background, along with his passion and initiative for learning, has led him to a profound understanding of power electronics. He manages the technical copywriting for programmable power and testing topics at ActionPower through product-marketing-driven approaches.